On-Chip Determination of C-Reactive Protein Using Magnetic Particles in Continuous Flow

Chayakom Phurimsak, Mark D. Tarn, Sally A. Peyman, John Greenman, Nicole Pamme*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

40 Citations (Scopus)

Abstract

We demonstrate the application of a multilaminar flow platform, in which functionalized magnetic particles are deflected through alternating laminar flow streams of reagents and washing solutions via an external magnet, for the rapid detection of the inflammatory biomarker, C-reactive protein (CRP). The two-step sandwich immunoassay was accomplished in less than 60 s, a vast improvement on the 80–300 min time frame required for enzyme-linked immunosorbent assays (ELISA) and the 50 min necessary for off-chip magnetic particle-based assays. The combination of continuous flow and a stationary magnet enables a degree of autonomy in the system, while a detection limit of 0.87 μg mL–1 makes it suitable for the determination of CRP concentrations in clinical diagnostics. Its applicability was further proven by assaying real human serum samples and comparing those results to values obtained using standard ELISA tests.
Original languageEnglish
Pages (from-to)10552-10559
Number of pages8
JournalAnalytical Chemistry
Volume86
Issue number21
DOIs
Publication statusPublished - 4 Nov 2014

Keywords

  • Assays
  • Carbon fiber
  • Enzyme-Linked Immunosorbent Assays
  • Fluorescence
  • Reagents

ASJC Scopus subject areas

  • Analytical Chemistry

Fingerprint

Dive into the research topics of 'On-Chip Determination of C-Reactive Protein Using Magnetic Particles in Continuous Flow'. Together they form a unique fingerprint.

Cite this