Dual-Band CW Radar System for Vital Signs Detection With Interference Cancellation Method

Chang Huan, Panagiota Kontou, Souheil Ben Smida, Dimitris E. Anagnostou

Research output: Contribution to journalArticlepeer-review

90 Downloads (Pure)

Abstract

A novel dual-band continuous-wave (CW) radar system is proposed. The radar is designed for vital signs detection in terms of breathing rate (BR), heart rate (HR), and radar cardiograph (RCG) waveforms. The system operates at two ISM band frequencies: 2.45 and 5.8 GHz. Two five-port receivers are used as IQ demodulators. A newly developed interference cancellation method is proposed. This new method increases the reliability of heartbeat detection by canceling the interfering respiration signal in terms of its fundamental frequency component. This method is first demonstrated in the frequency domain through signal processing and second implemented in the time domain using hardware. High detection accuracy is demonstrated for several experimental scenarios. The measurement results show that the proposed system detects vital signs effectively with a relative error of less than 5% at distances (between the radar and the measured subject) of up to 2.5 m using the frequency-domain method and 1.5 m using the time-domain method.
Original languageEnglish
Article number10759089
JournalIEEE Transactions on Microwave Theory and Techniques
Early online date20 Nov 2024
DOIs
Publication statusPublished - 20 Nov 2024

Keywords

  • Continuous-wave(CW) radar
  • dual band
  • heartbeat
  • interference suppression
  • respiration cancellation
  • vital signs detection

ASJC Scopus subject areas

  • Radiation
  • Condensed Matter Physics
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Dual-Band CW Radar System for Vital Signs Detection With Interference Cancellation Method'. Together they form a unique fingerprint.

Cite this